RecallWatchUSA

Home / Recalls / Medical Devices

TriMed Inc. recalls Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO;

Recalled April 18, 2018 · FDA recall Z-1373-2018 · TriMed

Hazard
Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy. · FDA Class II
Units
1525 devices
Sold at
US, China, Latin America

Description

Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy.

Product
Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; Hex Driver Part Number DVHX-2.5/130C AO; Hex Driver Part Number HXDRIVR-3.0 AO; Hex Driver Part Number HXDRIVER-3.5AO
Manufacturer(s)
TriMed Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

TriMed Inc. recalls Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; | RecallWatchUSA