TriMed Inc. recalls Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22

Recalled June 27, 2018 · FDA recall Z-2263-2018 · TriMed

Hazard
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized. · FDA Class II
Units
323
Sold at
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden

Description

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Product
Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22
Manufacturer(s)
TriMed Inc.