TriMed Inc. recalls Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10
Recalled June 27, 2018 · FDA recall Z-2265-2018 · TriMed
Hazard
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized. · FDA Class II
Units
1,194
Sold at
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
Description
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
Product
Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10; e) ES-15x12; f) ES-10x14; g) ES-10x15
Manufacturer(s)
TriMed Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →