RecallWatchUSA

Home / Recalls / Medical Devices

Trividia Health, Inc. recalls Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description

Recalled March 18, 2026 · FDA recall Z-1548-2026 · Trividia Health

Hazard
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose. · FDA Class I
Units
1,407,465
Sold at
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.

Description

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

Product
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE METRIX GO Kit/RF4031-01, Hyvee TRUE METRIX GO Kit/RF4048-01, Rite Aid TRUE METRIX GO Kit/RF4066-01, TopCo TRUE METRIX GO Kit/RF4209-01, TRUE METRIX GO Kit/RF4H01-01BK, TRUE METRIX GO NFRS Meter Only/RF4H01-40, TRUE METRIX GO Kit (Jamaica - mmol/L)/RF4i29-11BK, TRUE METRIX GO Kit (Australia - mmol/L)/RF4i81-11BK, TRUE METRIX GO Kit (United Kingdom - mmol/L)/RF4i82-11BK, TRUE METRIX GO Starter Kit (United Kingdom - mmol/L)/RF4i82-12BK
Manufacturer(s)
Trividia Health, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov