RecallWatchUSA

Home / Recalls / Medical Devices

Trumpf Medical Systems, Inc. recalls Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5

Recalled January 6, 2016 · FDA recall Z-0563-2016 · Trumpf Medical Systems

Hazard
The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall. · FDA Class II
Units
20,713
Sold at
USA (nationwide) Distribution to the states of : AL, AK, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KY, ME, MA, MN, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WY., and Internationally to Canada.

Description

The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.

Product
Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5
Manufacturer(s)
Trumpf Medical Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Trumpf Medical Systems, Inc. recalls Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5 | RecallWatchUSA