RecallWatchUSA

Home / Recalls / Medical Devices

Turncare, Inc recalls Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Vers

Recalled June 18, 2025 · FDA recall Z-1967-2025 · Turncare

Hazard
A software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy. · FDA Class II
Units
1081 units
Sold at
US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut

Description

A software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.

Product
Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-59 (Fix)
Manufacturer(s)
Turncare, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Turncare, Inc recalls Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Vers | RecallWatchUSA