Home / Recalls / Medical Devices
U&I CORP. recalls Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 -
Recalled May 6, 2020 · FDA recall Z-1823-2020 · U&I
Hazard
The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians. · FDA Class II
Units
42
Sold at
US Nationwide distribution in the state of CA.
Description
The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians.
Product
Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titanium markers.
Manufacturer(s)
U&I CORP.
Made in
Korea (the Republic of)
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →