Home / Recalls / Medical Devices
UFSK - International OSYS Gmbh recalls UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeut
Recalled July 26, 2023 · FDA recall Z-2212-2023 · UFSK - International OSYS
Hazard
Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position · FDA Class II
Units
239 chairs US
Sold at
Nationwide
Description
Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position
Product
UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeutic treatments of temporary to short-term duration, especially in ophthalmology Art.no. / REF no.: 15000004 Art.no. / REF no.: 15000006
Manufacturer(s)
UFSK - International OSYS Gmbh
Made in
Germany
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →