Home / Recalls / Medical Devices
UIH Technologies LLC recalls Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT ATLAS MD, Rx Only
Recalled October 29, 2025 · FDA recall Z-0221-2026 · UIH Technologies
Hazard
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement. · FDA Class II
Units
2 units
Sold at
US: OUS:
Description
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Product
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT ATLAS MD, Rx Only
Manufacturer(s)
UIH Technologies LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →