RecallWatchUSA

Home / Recalls / Medical Devices

UIH Technologies LLC recalls Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 78

Recalled October 29, 2025 · FDA recall Z-0224-2026 · UIH Technologies

Hazard
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement. · FDA Class II
Units
3 units
Sold at
US: OUS:

Description

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Product
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 780 MD, Rx Only
Manufacturer(s)
UIH Technologies LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov