RecallWatchUSA

Home / Recalls / Medical Devices

Ultra Clean Systems, Inc. recalls Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning

Recalled June 4, 2025 · FDA recall Z-1859-2025 · Ultra Clean Systems

Hazard
During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation. · FDA Class II
Units
96 systems
Sold at
Worldwide - U.S. Nationwide distribution including in the states of AZ, CA, CO, FL, ID, IL, IN, KS, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SD, TX, VA, WA, WI, and WY. The country of Canada.

Description

During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.

Product
Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system
Manufacturer(s)
Ultra Clean Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov