RecallWatchUSA

Home / Recalls / Medical Devices

Varian Medical Systems, Inc. Oncology Systems recalls Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Mo

Recalled December 26, 2012 · FDA recall Z-0531-2013 · Varian Medical Systems, Inc. Oncology Systems

Hazard
Varian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology [ARIA RO] DICOM import/export functionality where a 360 degree arc field may convert to static field during DICOM export or import. · FDA Class II
Units
264 units
Sold at
Worldwide Distribution

Description

Varian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology [ARIA RO] DICOM import/export functionality where a 360 degree arc field may convert to static field during DICOM export or import.

Product
Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Model Number: HIT, Affected Versions: ARIA version 11.0 below 11.0.55, Build 11.0.28 and 11.0.34, Reference/FSCA Identifier: CP-09798; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA
Manufacturer(s)
Varian Medical Systems, Inc. Oncology Systems

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Varian Medical Systems, Inc. Oncology Systems recalls Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Mo | RecallWatchUSA