RecallWatchUSA

Home / Recalls / Medical Devices

Varian Medical Systems, Inc. Oncology Systems recalls Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-0

Recalled October 3, 2012 · FDA recall Z-2462-2012 · Varian Medical Systems, Inc. Oncology Systems

Hazard
Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose. · FDA Class II
Units
6087
Sold at
Worldwide Distribution - USA (nationwide) and Internationally. . .

Description

Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose.

Product
Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-08922; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithim is specifically indicated for planning proton treatment of neoplasms of the eyes.
Manufacturer(s)
Varian Medical Systems, Inc. Oncology Systems

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Varian Medical Systems, Inc. Oncology Systems recalls Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-0 | RecallWatchUSA