Varian Medical Systems, Inc. Oncology Systems recalls Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algori
Recalled February 27, 2013 · FDA recall Z-0867-2013 · Varian Medical Systems, Inc. Oncology Systems
Hazard
The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used. · FDA Class II
Units
1168 units
Sold at
Worldwide Distribution - USA Nationwide
Description
The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
Product
Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Manufacturer(s)
Varian Medical Systems, Inc. Oncology Systems
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →