RecallWatchUSA

Home / Recalls / Medical Devices

Varian Medical Systems, Inc. recalls Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The Varian Head

Recalled October 19, 2016 · FDA recall Z-0077-2017 · Varian Medical Systems

Hazard
Varian Medical Systems has received a report that a user was able to easily rotate the head frame posts when attached to the mounting cam on the metal head ring. There was no report of serious injury due to this issue · FDA Class II
Units
75 devices
Sold at
US Distribution and Internationally to 1 site in Canada, India and Morocco

Description

Varian Medical Systems has received a report that a user was able to easily rotate the head frame posts when attached to the mounting cam on the metal head ring. There was no report of serious injury due to this issue

Product
Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The Varian Head Frame System is for use with a computed tomography scanner to perform imaging for treatment planning and a charged particle accelerator to perform immobilization of the treatment target for stereotactic radiosurgery or radiotherapy treatments on cranial lesions, tumors and conditions where radiation treatment is indicated.
Manufacturer(s)
Varian Medical Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov