Varian Medical Systems, Inc. recalls Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patien

Recalled December 24, 2014 · FDA recall Z-0526-2015 · Varian Medical Systems

Hazard
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module. · FDA Class II
Units
32 units
Sold at
Worldwide Distribution.

Description

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

Product
Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.
Manufacturer(s)
Varian Medical Systems, Inc.