Home / Recalls / Medical Devices
Ventana Medical Systems Inc recalls CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001, model 790-
Recalled September 26, 2018 · FDA recall Z-3006-2018 · Ventana Medical Systems
Hazard
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur. · FDA Class I
Units
1655 units
Sold at
worldwide
Description
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
Product
CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001, model 790-4296 Immunohistochemistry (IHC) for in vitro diagnostic use.
Manufacturer(s)
Ventana Medical Systems Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →