Verathon, Inc. recalls GlideScope Titanium Single-Use Video Laryngoscope. It is a single-use video laryngoscope system (GlideScope Titanium Sys

Recalled March 16, 2016 · FDA recall Z-1035-2016 · Verathon

Hazard
The firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when GlideScope Titanium SU laryngoscope blades are in use. Video "flickering" appears as the intermittent break-up of the on-screen video image, appearing as either distorted horizontal or vertical bars of displaced video signal. · FDA Class I
Units
7733 units (6377 units in the US and 1356 units International)
Sold at
Worldwide Distribution - US (nationwide) and to countries of: Australia, Belgium, Canada, Denmark, France, Germany, Hong Kong, Italy, Luxembourg, Saudi Arabia, Singapore, Spain, Sweden, The Netherlands, U.A.E., and United Kingdom.

Description

The firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when GlideScope Titanium SU laryngoscope blades are in use. Video "flickering" appears as the intermittent break-up of the on-screen video image, appearing as either distorted horizontal or vertical bars of displaced video signal.

Product
GlideScope Titanium Single-Use Video Laryngoscope. It is a single-use video laryngoscope system (GlideScope Titanium System) consists of a sterile-packaged, single-use video laryngoscope, multi-use Smart Cable and portable GlideScope video monitor. The system incorporates a high-resolution, full-color digital camera and monitor for real-time viewing and recording. Affected Product Names and Part Numbers: -LoPro S3: 0574-0130 (single blade) or 0270-0769 (box of 10) -LoPro S4: 0574-0131 (single blade) or 0270-0770 (box of 10) -MAC S3: 0574-0132 (single blade) or 0270-0771 (box of 10) -MAC S4: 0574-0133 (single blade) or 0270-0772 (box of 10) The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures.
Manufacturer(s)
Verathon, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Verathon, Inc. recalls GlideScope Titanium Single-Use Video Laryngoscope. It is a single-use video laryngoscope system (GlideScope Titanium Sys | RecallWatchUSA