Verathon, Inc. recalls GlideScope Video Laryngoscope Reusable Blades (GVL 3, 4, 5 and AVL 2, 3, 4, 5). GVL: The GlideScope Video Laryngoscop
Recalled February 24, 2016 · FDA recall Z-0811-2016 · Verathon
Hazard
Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such that exceeding this number of cycles in the specific chemical agent may affect the potential life of the product. · FDA Class II
Units
total 9088 units ( 5786 units in the US and 3302 units International)
Sold at
Worldwide Distribution.
Description
Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such that exceeding this number of cycles in the specific chemical agent may affect the potential life of the product.
Product
GlideScope Video Laryngoscope Reusable Blades (GVL 3, 4, 5 and AVL 2, 3, 4, 5). GVL: The GlideScope Video Laryngoscope is a video laryngoscope that incorporates a high-resolution color camera, an LED light source, a rechargeable lithium battery, and video output to a remote monitor. It consists of a video laryngoscope and a portable video monitor. AVL: The GlideScope AVL Video Monitor system has the same intended use as the GVL System. The difference between systems is that the AVL has a digital platform and the GVL an analog platform.
Manufacturer(s)
Verathon, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →