Viewray Incorporated recalls MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian¿ Sys
Recalled April 6, 2016 · FDA recall Z-1309-2016 · Viewray Incorporated
Hazard
When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a result there is the potential to deliver dose to the initial isocenter rather than the new location. · FDA Class II
Units
5 units
Sold at
Worldwide Distribution - US including CA, FL, MO & WI., and International distribution to the Republic of Korea and The Netherlands.
Description
When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a result there is the potential to deliver dose to the initial isocenter rather than the new location.
Product
MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian¿ System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated