Villa Sistemi Medicali S.P.A. recalls Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray syst
Recalled January 8, 2020 · FDA recall Z-0764-2020 · Villa Sistemi Medicali S.P.A
Hazard
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years. · FDA Class II
Units
3
Sold at
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.
Description
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
Product
Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray system suitable for radiography and fluoroscopy examinations. The system is stationary mounted. The movements of the table and the tube stand are motorized and remotely controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
Manufacturer(s)
Villa Sistemi Medicali S.P.A.
Made in
Italy