RecallWatchUSA

Home / Recalls / Medical Devices

Visicu, Inc. recalls The eCareManager system. Software intended for use in data collection, storage and clinical information management with

Recalled December 16, 2015 · FDA recall Z-0399-2016 · Visicu

Hazard
A software defect may cause incorrect medication order change. If the user decides to edit the order prior to electronic signature AND selects the return to Previous Screen function, the previously selected medication is replaced with whatever medication is in the number one spot on the complete medication list. At the stage where the order is signed, the medication displayed may be incorrect. · FDA Class II
Units
43
Sold at
US distribution only.

Description

A software defect may cause incorrect medication order change. If the user decides to edit the order prior to electronic signature AND selects the return to Previous Screen function, the previously selected medication is replaced with whatever medication is in the number one spot on the complete medication list. At the stage where the order is signed, the medication displayed may be incorrect.

Product
The eCareManager system. Software intended for use in data collection, storage and clinical information management with independent bedside devices, and ancillary systems that are connected either directly or through networks. For use in a hospital environment. eCareManager allows clinicians to generate a PDF display of an order (new, change, discontinue), which is printed at the bedside and/or hospital pharmacy. This printed PDF should be reviewed by the clinician and verified by the hospital pharmacist and entered into the hospital pharmacy system.
Manufacturer(s)
Visicu, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov