Volcano Corp recalls ComboMap Intravascular Flow and Pressure System; Product/Part number: Combo Map - 802745-001. Intravascular Imag
Recalled July 15, 2015 · FDA recall Z-2045-2015 · Volcano
Hazard
A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output. · FDA Class II
Units
227 total units - all models
Sold at
Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.
Description
A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
Product
ComboMap Intravascular Flow and Pressure System; Product/Part number: Combo Map - 802745-001. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.
Manufacturer(s)
Volcano Corp
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →