RecallWatchUSA

Home / Recalls / Medical Devices

Volcano Corp recalls Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 8

Recalled August 6, 2025 · FDA recall Z-2221-2025 · Volcano

Hazard
Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire" · FDA Class II
Units
233,817
Sold at
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of BR, PL, SK, IT, DK, DE, GB, AT, KW, CR, ES, NL, LV, MT, IE, CH, ZA, RO, FI, PR, FR, CA, PA, IL, SE, PT, HR, GR, NZ, HK, TW, MY, TH, MX, AR, PE, AE, BG, AU, JP, ID, RS, CO, SG, BE.

Description

Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Product
Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.
Manufacturer(s)
Volcano Corp

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Volcano Corp recalls Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 8 | RecallWatchUSA