RecallWatchUSA

Home / Recalls / Medical Devices

Waldemar Link GmbH & Co. KG (Mfg Site) recalls LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 19-Indicated for patients with Mobility-limiting diseases, frac

Recalled August 16, 2023 · FDA recall Z-2350-2023 · Waldemar Link GmbH & Co. KG (Mfg Site)

Hazard
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue · FDA Class II
Units
24 units
Sold at
US Nationwide distribution.

Description

Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue

Product
LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 19-Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-119/00
Manufacturer(s)
Waldemar Link GmbH & Co. KG (Mfg Site)
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Waldemar Link GmbH & Co. KG (Mfg Site) recalls LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 19-Indicated for patients with Mobility-limiting diseases, frac | RecallWatchUSA