Home / Recalls / Medical Devices
WASATCH PHOTONICS recalls Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP,
Recalled September 17, 2025 · FDA recall Z-2478-2025 · WASATCH PHOTONICS
Hazard
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration. · FDA Class II
Units
192
Sold at
U.S.
Description
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
Product
Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.
Manufacturer(s)
WASATCH PHOTONICS
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →