WELCH ALLYN, INC/MORTARA recalls H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, P

Recalled May 6, 2020 · FDA recall Z-1812-2020 · WELCH ALLYN, INC/MORTARA

Hazard
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment. · FDA Class II
Units
78,521 units total
Sold at
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Description

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Product
H12+ and patient cable for the H12+ Holter Recorders with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP. The 24-hour H12+ recorder uses a single AA alkaline battery to provide continuous 12-lead data recorded over a 24-hour period and a removable 24-hour compact flash (CF) card for data storage.
Manufacturer(s)
WELCH ALLYN, INC/MORTARA

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov