RecallWatchUSA

Home / Recalls / Medical Devices

WELCH ALLYN, INC/MORTARA recalls Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA

Recalled May 6, 2020 · FDA recall Z-1816-2020 · WELCH ALLYN, INC/MORTARA

Hazard
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment. · FDA Class II
Units
78,521 units total
Sold at
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Description

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Product
Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for continuous monitoring in either stationary or portable applications.
Manufacturer(s)
WELCH ALLYN, INC/MORTARA

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

WELCH ALLYN, INC/MORTARA recalls Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA | RecallWatchUSA