Home / Recalls / Medical Devices
Western/Scott Fetzer Company recalls Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Dig
Recalled April 19, 2023 · FDA recall Z-1376-2023 · Western/Scott Fetzer
Hazard
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow. · FDA Class II
Units
24,287 units
Sold at
US Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.
Description
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
Product
Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Digital VIPR System (with E-size cylinder) MNDR-600, Western Oxytote DTE Digital VIPR (head only) MNDS-602, Western Oxytote DTE Digital VIPR System (with D-size cylinder) MNDS-603, Western Oxytote DTE Digital VIPR System (with E-size cylinder)
Manufacturer(s)
Western/Scott Fetzer Company
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →