Home / Recalls / Medical Devices
Westmed, Inc. recalls REF 0562, Adult Elongated O2/CO2 Mask, 10'X0.50 GSL w/ Swivel Elbow, 2" Pigtail w/Female Luer Connectors, 7' Oxygen Supp
Recalled January 6, 2021 · FDA recall Z-0746-2021 · Westmed
Hazard
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment. · FDA Class II
Units
6000 units
Sold at
Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom
Description
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
Product
REF 0562, Adult Elongated O2/CO2 Mask, 10'X0.50 GSL w/ Swivel Elbow, 2" Pigtail w/Female Luer Connectors, 7' Oxygen Supply Tubing, Rx Only, CE0482, (01)00709078001697 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.
Manufacturer(s)
Westmed, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →