Westmed, Inc. recalls Westmed DISPOSABLE RESPIRATORY PRODUCTS NON-REBREATHER MASK, REF 3228-e, RX Only, CE0482, (01)10709078008020 - Product

Recalled January 6, 2021 · FDA recall Z-0742-2021 · Westmed

Hazard
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment. · FDA Class II
Units
37,300 units
Sold at
Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom

Description

There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.

Product
Westmed DISPOSABLE RESPIRATORY PRODUCTS NON-REBREATHER MASK, REF 3228-e, RX Only, CE0482, (01)10709078008020 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.
Manufacturer(s)
Westmed, Inc.