WillowWood Global Llc recalls Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a

Recalled May 26, 2021 · FDA recall Z-1594-2021 · WillowWood Global

Hazard
the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension. · FDA Class II
Units
815 parts
Sold at
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, KY, LA, MA, ME, MN, MO, MS, NJ, NM, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and the countries of Germany, Netherlands

Description

the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.

Product
Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha Lanyard - 700-AIS101, Alpha Interlock Suspension System with Alpha Lanyard w/Titanium Pyramid - 700-AIS107, Alpha Interlock Suspension System Lanyard Body ONLY
Manufacturer(s)
WillowWood Global Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

WillowWood Global Llc recalls Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a | RecallWatchUSA