Wilson-Cook Medical Inc. recalls Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA
Recalled February 25, 2026 · FDA recall Z-1362-2026 · Wilson-Cook Medical
Hazard
Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting. · FDA Class II
Units
153
Sold at
U.S. Distribution to states of: CA, CO, IN, and NC.
Description
Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
Product
Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also aids in bridging difficult strictures during ERCP. It is indicated for sphincterotome-aided, wire-guided selective cannulation of the biliary ducts.
Manufacturer(s)
Wilson-Cook Medical Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →