Home / Recalls / Medical Devices
Wipro GE Healthcare Private Ltd. recalls GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148;
Recalled October 2, 2024 · FDA recall Z-3081-2024 · Wipro GE Healthcare Private
Hazard
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits. · FDA Class II
Units
137 units
Sold at
US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.
Description
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
Product
GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
Manufacturer(s)
Wipro GE Healthcare Private Ltd.
Made in
India
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →