RecallWatchUSA

Home / Recalls / Medical Devices

WISSNER-BOSSERHOFF GMBH recalls SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-01

Recalled November 20, 2024 · FDA recall Z-0329-2025 · WISSNER-BOSSERHOFF

Hazard
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved. · FDA Class II
Units
77 units
Sold at
US Nationwide distribution in the states of NC.

Description

When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.

Product
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Manufacturer(s)
WISSNER-BOSSERHOFF GMBH
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov