Home / Recalls / Medical Devices
Wright Medical Technology, Inc. recalls Wright EVOLVE TRIAD Plate Cutter, REF 49510120
Recalled July 20, 2022 · FDA recall Z-1366-2022 · Wright Medical Technology
Hazard
The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use · FDA Class II
Units
231
Sold at
Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico
Description
The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use
Product
Wright EVOLVE TRIAD Plate Cutter, REF 49510120
Manufacturer(s)
Wright Medical Technology, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →