RecallWatchUSA

Home / Recalls / Medical Devices

Ziehm Imaging Inc recalls Ziehm Vision RFD, Mobile C-arm for Mobile Fluoroscopic Imaging

Recalled August 31, 2016 · FDA recall Z-2641-2016 · Ziehm Imaging

Hazard
Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted. · FDA Class II
Units
2 devices
Sold at
US Distribution including Puerto Rico and to the states of :FL, GA, VA,MA. ME, MI, NY, NJ, PA, RI, IL, TX, AZ, CA, NV, UT, and WA.

Description

Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.

Product
Ziehm Vision RFD, Mobile C-arm for Mobile Fluoroscopic Imaging
Manufacturer(s)
Ziehm Imaging Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov