RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer Biomet, Inc. recalls Bone Dowell Harvest Tubes

Recalled May 17, 2017 · FDA recall Z-2059-2017 · Zimmer Biomet

Hazard
mislabeled as 8mm tube but it is actually 9 mm in size. A 9mm subcomponent was erroneously substituted and etched as 8mm for the actual 8mm subcomponent part. · FDA Class II
Units
49
Sold at
No US distribution. JAPAN NETHERLANDS SINGAPORE SWITZERLAND

Description

mislabeled as 8mm tube but it is actually 9 mm in size. A 9mm subcomponent was erroneously substituted and etched as 8mm for the actual 8mm subcomponent part.

Product
Bone Dowell Harvest Tubes
Manufacturer(s)
Zimmer Biomet, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov