Zimmer Biomet, Inc. recalls Drill Bit Quick-Connect 2.7 mm Diameter 100 mm Length, Item Number/EDI 00480610027, Nonsterile. orthopedic surgical inst

Recalled October 4, 2017 · FDA recall Z-3194-2017 · Zimmer Biomet

Hazard
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture. · FDA Class II
Units
701 units
Sold at
Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.

Description

Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.

Product
Drill Bit Quick-Connect 2.7 mm Diameter 100 mm Length, Item Number/EDI 00480610027, Nonsterile. orthopedic surgical instrument.
Manufacturer(s)
Zimmer Biomet, Inc.