RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer Biomet, Inc. recalls ECT Internal Fracture Fixation Drill 2.5 mm Diameter (.098 Inch), Item Number/EDI 00231802500, Nonsterile. orthopedic su

Recalled October 4, 2017 · FDA recall Z-3166-2017 · Zimmer Biomet

Hazard
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture. · FDA Class II
Units
3,232 units
Sold at
Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.

Description

Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.

Product
ECT Internal Fracture Fixation Drill 2.5 mm Diameter (.098 Inch), Item Number/EDI 00231802500, Nonsterile. orthopedic surgical instrument.
Manufacturer(s)
Zimmer Biomet, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov