RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer Biomet, Inc. recalls Persona Partial Knee Spacer Blocks To assist in partial knee arthroplasty surgery

Recalled April 4, 2018 · FDA recall Z-1240-2018 · Zimmer Biomet

Hazard
Zimmer Biomet is conducting a medical device recall for the Persona Partial Knee Spacer Blocks in sizes 8mm, 9mm, 10mm, 12mm, and 14mm. Complaints received indicated that flexion has the potential to feel tight during range-of-motion trialing and that the final trial has the potential to appear looser in extension. · FDA Class II
Units
430
Sold at
US Nationwide Distribution.

Description

Zimmer Biomet is conducting a medical device recall for the Persona Partial Knee Spacer Blocks in sizes 8mm, 9mm, 10mm, 12mm, and 14mm. Complaints received indicated that flexion has the potential to feel tight during range-of-motion trialing and that the final trial has the potential to appear looser in extension.

Product
Persona Partial Knee Spacer Blocks To assist in partial knee arthroplasty surgery
Manufacturer(s)
Zimmer Biomet, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Zimmer Biomet, Inc. recalls Persona Partial Knee Spacer Blocks To assist in partial knee arthroplasty surgery | RecallWatchUSA