RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer Biomet, Inc. recalls Segmental fluted stem, 17x190mm str, Sterile, Knee joint femorotibial meta/polymer constrained cemented prosthesis- Sp

Recalled June 21, 2017 · FDA recall Z-2568-2017 · Zimmer Biomet

Hazard
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures. · FDA Class II
Units
9867
Sold at
Worldwide Distribution - US (nationwide) Foreign: Argentina, Australia, Belgium, Brazil, Bridgend Wales, Canada, Chile, Colombia, Dominican Republic, England, Eschbach, France, India, Italy, Japan, Jordan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Panama, Peru, Russia, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, Thailand

Description

Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.

Product
Segmental fluted stem, 17x190mm str, Sterile, Knee joint femorotibial meta/polymer constrained cemented prosthesis- Specific Knee, Hip and Nail Implants
Manufacturer(s)
Zimmer Biomet, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov