Home / Recalls / Medical Devices
Zimmer Biomet, Inc. recalls Zimmer Natural Nail¿ System - Tear Drop Guide Wire, 3.0 mm Diameter, 70 cm Length Item Number: 47-2490-098-00 Produ
Recalled December 12, 2018 · FDA recall Z-0540-2019 · Zimmer Biomet
Hazard
Various Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility · FDA Class II
Units
12724
Sold at
Worldwide - US Nationwide Distribution Foreign: CANADA AUSTRALIA BRAZIL CHILE HONG KONG JAPAN MALAYSIA NETHERLANDS NICARAGUA SINGAPORE TAIWAN
Description
Various Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility
Product
Zimmer Natural Nail¿ System - Tear Drop Guide Wire, 3.0 mm Diameter, 70 cm Length Item Number: 47-2490-098-00 Product Usage: A single use surgical instrument used during orthopaedic surgery for implantation of a prosthesis.
Manufacturer(s)
Zimmer Biomet, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →