Zimmer Biomet Spine Inc. recalls Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number
Recalled June 14, 2017 · FDA recall Z-2318-2017 · Zimmer Biomet Spine
Hazard
Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification. · FDA Class II
Units
497 units
Sold at
Nationwide, military/VA/govt and foreign consignee (No Canadian consignees)
Description
Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.
Product
Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017
Manufacturer(s)
Zimmer Biomet Spine Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →