Home / Recalls / Medical Devices
Zimmer, Inc. recalls 00595001201 Femur CR-FLEX PCT FEM B-L Rx, Sterile; 00595001202 Femur CR-FLEX PCT FEM B-R
Recalled August 28, 2013 · FDA recall Z-2006-2013 · Zimmer
Hazard
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d · FDA Class II
Units
192,355
Sold at
Worldwide Distribution-USA (nationwide) and the country of Japan.
Description
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Product
00595001201 Femur CR-FLEX PCT FEM B-L Rx, Sterile; 00595001202 Femur CR-FLEX PCT FEM B-R Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Manufacturer(s)
Zimmer, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →