RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer, Inc. recalls 00780303601 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780303602 MIS 12/14 Slotted Femoral Head Provisional +

Recalled November 28, 2012 · FDA recall Z-0399-2013 · Zimmer

Hazard
Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat · FDA Class II
Units
56,516 all lots, all sizes
Sold at
Worldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.

Description

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat

Product
00780303601 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780303602 MIS 12/14 Slotted Femoral Head Provisional +0X36 00780303603 MIS 12/14 Slotted Femoral Head Provisional +3.5X36 00780303604 MIS 12/14 Slotted Femoral Head Provisional +7X36 00780303605 MIS 12/14 Slotted Femoral Head Provisional +10.5X36 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Manufacturer(s)
Zimmer, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Zimmer, Inc. recalls 00780303601 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780303602 MIS 12/14 Slotted Femoral Head Provisional + | RecallWatchUSA