Zimmer, Inc. recalls CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in total hip
Recalled April 5, 2023 · FDA recall Z-1294-2023 · Zimmer
Hazard
Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009 · FDA Class II
Units
1,168 untits
Sold at
AL AZ CO FL LA MI OH OK TX Foreign: Canada
Description
Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009
Product
CoCr Femoral Head, XS, 38/-8, Taper 12/14-Intended to be used as a modular head component for articulation in total hip arthroplasty. Item Number: 01.01012.384
Manufacturer(s)
Zimmer, Inc.