RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer, Inc. recalls Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279

Recalled February 11, 2015 · FDA recall Z-1052-2015 · Zimmer

Hazard
Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22. · FDA Class II
Units
67963
Sold at
Worldwide Distribution - USA Nationwide including DC, and PR and the countries of Australia, Austria, Belgium, Canada, Czech Republic, Dominican Republic, France, Germany, India, Iran, Israel, Italy, Japan, Luxembourg, Netherlands, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

Description

Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.

Product
Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The Persona TASPs are sterilizable instruments Intended for multiple uses. The TASP components are used to aid the surgeon in assessing the thickness of the articular surface required. The TASP construct is utilized while conducting a trial range of motion. The TASP consists of 3 parts: a TASP bottom, a TASP shim, and a TASP top. The shims come in varying thicknesses to allow trialing for specific Articular Surface implant thicknesses.
Manufacturer(s)
Zimmer, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov