Zimmer, Inc. recalls Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devi

Recalled October 14, 2015 · FDA recall Z-0027-2016 · Zimmer

Hazard
It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat · FDA Class II
Units
18 units
Sold at
Distributed in NC, AL, KS, TX, UT, CA, and SINGAPORE.

Description

It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat

Product
Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices.
Manufacturer(s)
Zimmer, Inc.