RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer, Inc. recalls ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and sta

Recalled May 8, 2024 · FDA recall Z-1723-2024 · Zimmer

Hazard
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention · FDA Class II
Sold at
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.

Description

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Product
ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-017-06
Manufacturer(s)
Zimmer, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov